FL-0001 specifications

Product information for FL-0001
Product codeFL-0001
Weight4 kg
DimensionsCylindrical; outside diameter 200 mm. Height excluding fittings: 20 mm. Height including fittings: 85 mm
Test material surface area177 cm² (circular)
Test chamber volume35 cm³
Air exchange rate100–1000 per hour
Air velocity over test material0.35–1.75 cm/s
Test chamber surface area180 cm²
Air inlet flow100–500 ml/min
Air outlet flow to sample tubes80–400 ml/min
Air outlet flow to vent20–100 ml/min
Material in contact with the sample streamCertified solid, acid-resisting, polished stainless steel (all internal contact surfaces are mirror-polished to minimize sink effects).
Connection to air supplyTwo 1/4″ PFA tubes, Swagelok Teflon cone
Connection for sample tubesTwo 1/4″ sample tubes can be connected for simultaneous sampling

System requirements

System requirements for FL-0001
Air supplyThe FLEC Cell must be supplied with well-controlled, clean and humidified air, e.g. FLEC Air Control FL-1000
Suction pumpsPrecision suction pumps for sample tubes can be used

Select airflow and sampling conditions according to the applicable method and your material. Contact CHEMATEC for configuration guidance.

How the measurement works

From surface emissions to laboratory results

The material closes the underside of the cell and defines the exposed surface. Controlled air carries released compounds to suitable sampling media. Analysis of the collected sample establishes which compounds were present and their concentrations.

  • Prepare the specimenUse the appropriate holder and document the exposed area, material age and conditioning.
  • Verify the flowsCheck the inlet flow and each sampling flow. Keep an excess-air path through the vent.
  • Choose the analytical routeAgree sampling media, duration and analysis with the laboratory for the compounds of interest.

For VOCs, ISO 16000-6 describes sorbent-tube sampling followed by thermal desorption and gas chromatography. ISO 16000-3:2026 covers active air sampling and analysis for formaldehyde and other carbonyl compounds. For acrolein, Annex A uses sorbent-tube sampling with thermal desorption and GC-MS because the DNPH method is not considered reliable for quantitative results. ISO 16000-3 addresses sampling and analysis; the emission-cell test method is covered by ISO 16000-10. Compare results using a consistent preparation, climate and sampling procedure.